Maintenance / Sterilization Procedures (Autoclaves Only)
Clean External Surface: Remove bur from handpiece and scrub with a toothbrush or 2" X 2" gauze using warm tap water. DO NOT IMMERSE HANDPIECE.
Dry: Thoroughly dry handpiece using gauze, paper towel or air syringe.
Clean/Lubricate Internal Surface: Using a combination cleaner/lubricant or DCL 90, spray handpiece in drive air hole and in chuck.
Expel Cleaner/Lubricant: Reinsert bur into handpiece head, connect handpiece to tubing and run for 5 seconds to thoroughly expel debris and excess lubricant.
The Handpiece Must Be Operated Only with A Bur. Do Not Operate Empty.
Clean Fiberoptic Bundle: Using a cotton swab with isopropyl alcohol, wipe the surface on both ends of handpiece.
Bag and Cycle in Autoclave: Place handpiece into autoclave bag or pouch. Cycle as per autoclave manufacturer's instruction. DO NOT EXCEED 275 Degrees F ( 135 C ).
Cool Down and Lubricate: Allow handpiece to return to room temperature. Lubricate handpiece as per instructions listed above. Expel excess lubricant as per previous instructions.